e-breathe Respiratory Protection Test Torso – for Standard Testing, Research, and Development

The e-breathe respiratory protection test torso is used in combination with a test head—either the EN-standard Sheffield head or ISO test heads in sizes S, SW, M, LN, and L. This configuration enables the testing of respiratory protection systems with an air supply and facepiece under realistic wearing conditions, such as self-contained breathing apparatus with full-face masks or powered air-purifying respirators with hoods.

The e-breathe respiratory protection test torso allows the direct mounting of the air supply unit on the body and ensures a secure connection of the respiratory interface to the test head. Via standardized interfaces, the torso can be connected to breathing machines or lung simulators, providing reproducible, realistic, and application-oriented test conditions for professional testing, research, and development.

Product Features
e-breathe Respiratory Protection Test Torso

Connections & Sensor Technology

  • Direct connection to the test head via measurement ports in the mouth and trachea area (CO₂ and pressure measurement)
  • EN 148-1 round threaded connection at the front and rear for integration with test equipment

Material & Construction

  • Robust, modular design consisting of five segments, tightly fastened and easy to disassemble for maintenance or expansion
  • Front and rear sections made of durable plastic
  • Front access opening for quick access to hoses and valves
  • Stable base, designed to support heavy respiratory protection equipment
  • Secure positioning of backpack harnesses and self-contained breathing apparatus, preventing slippage

Optional Head Movement Module

  • Automated head module simulating left and right head rotation
  • Enables the execution of two of the five exercises required for leakage testing according to ISO 16900-1 / EN 13274-1

Leakage & Tightness Testing

  • Completely leak-tight construction 
  • Suitable for testing in leakage chambers to determine inward leakage and to assess the protection class of respiratory protection systems

Technical Highlights

Areas of Application

Standards & Guidelines

Suitable Accessories